Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 6-7 REF:402567 component Lantern Knee System

FDA Recall Z-1630-2026 · Posted 2026-03-24

Recalling FirmOrthalign, Inc
Recall StatusOpen, Classified
Product CodeOLO
Date Initiated2026-02-10
Product Quantity7 units

Reason for Recall

Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm)

Firm Action

On February 10, 2026, OrthAlign issued a Urgent: Medical device Recall Notification via overnight mail. OrthAlign asked consignees to take the following actions: Please review your inventory for the affected products listed in Table 1. 1. Immediately remove and quarantine the affected products until the products are returned to OrthAlign. 2. Please share this recall notification with all members in your network as well as those where the affected devices have been transferred. 3. If you have further distributed the affected products, please identify and notify consignees. You may provide consignees a copy of this notification letter. 4. Please complete the attached response form within Five (5) business days and email the completed form to : [email protected] 5. Adverse actions are quality issues and can be reported to the Food and Drug Administration s MedWatch program: o Online: By completing and submitting the report online at www.fda.gov/medwatch o Regular mail or fax): download the form from www.fda.gov/MedWatch/getforms.htm or call 800-332-1088 to request a reporting form; then complete the form and mail it to the address on the pre-addressed form or submit by fax to 800-332-0178.

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide distribution in the state of UT.