Leica Wireless Footswitch, Type B used in conjunction with Leica Surgical Microscopes - Models M620, M844/820, M525 and M720; There are two wireless models: a) MFS20, stock number 10448406, a 12 function wireless footswitch type B with cross pedals, XY-actuator 10717602 and screw 10630134 for dust protection enclosed; b) MFS21, stock number 10448407, a 16 function wireless footswitch type B with cross pedals, XY-actuator 10717602 and screw 10630134 for dust protection enclosed;
FDA Recall Z-1633-2010 · Posted 2010-05-20
Advertisement
Recalling FirmLeica Microsystems, Inc.
Recall StatusTerminated
Product CodeFSO
Date Initiated2010-03-29
Date Terminated2011-02-15
Product Quantity3 units
Reason for Recall
The pedals of the Footswitch Type B stick and the Zoom/Focus moves to the end position.
Firm Action
Leica Microsystems sent Medical Device Correction letters dated 3/29/10 to the affected customers via first class mail on the same date, informing them of the potential for injury to the patient's eye if the cross pedal of the footswitch sticks. They were informed that their local Leica Microsystems representative will contact them to arrange for the replacement of the cross pedal. The letters also provided the interim necessary steps required to avoid the risk of patient injury while they are awaiting the cross pedal replacement, i.e. do not use the affected footswitches with a surgical microscope that has a BIOM accessory. The accounts were requested to complete the attached Acknowledgement Form, listing the serial numbers of the units at their facility and fax it back to Leica at 1-847-607-3147. A company contact was provided in the letter for questions.
Root Cause
Other
Distribution Pattern
Worldwide distribution -- U.S. (nationwide), Canada, Australia, Europe and Asia
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement