NAVIO Soft Tissue Protector, included in instrument kits. The NAVIO Tissue Protector is a reusable instrument designed to, when used as instructed, provide the required bone pin spacing for the Navio bone clamps and potentially reduce the amount of patient tissue affected when placing the bone pins.

FDA Recall Z-1634-2020 · Posted None

Recalling FirmSmith & Nephew, Inc.
Recall StatusTerminated
Product CodeOLO
Date Initiated2020-02-27
Date Terminated2024-08-02
Product Quantity348 units

Reason for Recall

Potential for the NAVIO Soft Tissue Protector to become stuck or bound to the bone pin intraoperatively when inserting the NAVIO bone pins.

Firm Action

The firm, Smith & Nephew, notified customers on about 03/03/2020 via "Urgent Medical Device Recall Notice" letter. The customers were Instructed to examine inventory and quarantine affected devices, notify customers if further distributed, and complete and return the Response Form to [email protected] or fax to 1-901-566-7975. Questions and concerns can be directed to EieldActions@smlth-nephew .com. or 901-399-1970.

Root Cause

Device Design

Distribution Pattern

Nationwide distribution to AL, AZ, CA, CT, FL, GA, HI, IA, IL, KS, LA, MA, MI, MN, MS, MT, NC, NJ, NY, OH, OK, PA, TN, TX, VA, WA, WI, WV. International distribution to Turkey, United Kingdom, Thailand, Switzerland, France, Australia.