Terminated

TRIGEN INTERTAN NAIL: (1) REF 71675201, 10 MM X 18 CM, 125 degree, (2) REF 71675202, 11.5 MM X 18 CM, 125 degree, (3) REF 71675203, 13 MM X 18 CM, 125 degree, (4) REF 71675204, 10 MM X 20 CM, 125 degree, (5) REF 71675205, 11.5 MM X 20 CM, 125 degree, (6) REF 71675207, 10 MM X 18 CM, 130 degree, (7) REF 71675208, 11.5 MM X 18 CM, 130 degree, (8) REF 71675211, 11.5 MM X 20 CM, 130 degree, Smith & Nephew, Inc. Orthopedic fixation device.

FDA Recall Z-1636-2014 · Posted 2014-05-20

Recalling FirmSmith & Nephew Inc
Recall StatusTerminated
Product CodeJDS
Date Initiated2014-04-04
Date Terminated2017-07-01
Product Quantity1,253 units total

Reason for Recall

A manufacturing error resulted in the set screw being inserted too deep in the affected nails.

Firm Action

The firm initiated the recall via an "Urgent - Product Recall 1st Notification - Urgent" letter and e-mail on 04/04/2014, identifying the product, explaining the problem, and the potential risk associated with using the product. Customers are to inspect their inventory, and locate and quarantine any affected product. Additionally, the Inventory Return Certification form should be completed and returned along with any affected product identified, once a Return Authorization number has been obtained from Smith & Nephew's Safety Affairs Department.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution -- USA, France, Germany, Sweden, Switzerland, UK, Canada, Australia, Japan, China, India, Dubai, South Africa, South Korea, Mexico, and Venezuela.