AmniSure ROM Test, Material Nos. FMRT-1-25-US (US Distribution, 25 tests/box), FMRT-1-10-US (US Distribution, 10 tests/box), FMRT-1-25-CA (Canadian Distribution; 25 tests/box)
FDA Recall Z-1636-2020 · Posted None
Advertisement
Recalling FirmQiagen Sciences LLC
Recall StatusTerminated
Product CodeNQM
Date Initiated2020-02-21
Date Terminated2020-10-05
Product Quantity1654 kit boxes
Reason for Recall
An image on the associated polybag contains an incorrect rinse time of 30 (thirty seconds). The correct rinse time is 1 (one) minute.
Firm Action
On February 21, 2020, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed of the incorrect rinse time on the product's labeling, and reminded of the correct rinse time of 1 (one) minute.
Customers were asked to immediately share the contents of the customer letter with all users of the AmniSure ROM Test in their facility to ensure awareness. Test users should refer to the instructions for use (IFU) included in each kit box. The firm is not asking for product to be returned. Customers were asked to complete and return an attached response form.
Root Cause
Employee error
Distribution Pattern
Nationwide domestic distribution. Foreign distribution to Canada.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.