BD Beaver Arthro-Lok Pointed Tip 4mm Blade, REF 376780. Becton Dickinson and Company, Waltham, MA 02454. Intended for use in arthroscopic procedures.
FDA Recall Z-1638-2010 · Posted 2010-05-21
Advertisement
Recalling FirmBecton Dickinson and Company
Recall StatusTerminated
Product CodeHTS
Date Initiated2009-06-12
Date Terminated2012-01-26
Product Quantity354 units
Reason for Recall
Mislabeled: Arthro-Lok Pointed Tip 4mm Blade curving right instead of left.
Firm Action
Becton Dickinson issued an "Urgent: Field Safety Notice" beginning June 25, 2009 by telephone and letter fax to remove and discard affected product from their possession. Customer verification responses were requested.
For further information, contact Becton Dickinson and Company at 1-781-906-7900.
Root Cause
Packaging process control
Distribution Pattern
Worldwide Distribution -- United States, Japan and Belgium.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.