Terminated

- Calf Garment Standard Ref. L501-M Lot 122549, 122985 - Calf Garment, DVT10 Standard Ref. DVT10 Lot 122570, 12546 - Calf Garment, DVT20 Large Ref. DVT20 Lot 122549, 122570 - SCD Express Compression Sleeve, Knee Length, (White) Large Ref. 9789 Lot 122574 - SCD Express Compression Sleeve, Knee Length, (White) Medium Ref. 9529 Lot 122547, 122574 - SCD Express Compression Sleeve, Thigh Length, (White) Medium Ref. 9530 Lot 122574 Compressible limb sleeve. - Thigh Garment DVT30 Medium Ref. DVT30 Lot 122549, 122985 - Thigh Garment DVT40 Large Ref. DVT40 Lot 122549, 122985 - Vaso-Force Intermittent DVT Calf Garment, Large, Calf Sizes up to 24" 24" Ref. MDS601L Lot 122982, 122986 - Vaso-Force Intermittent DVT Calf Garment, Medium, Calf Sizes up to 18" 18" Ref. MDS601M Lot 122986, 122984 - VENAFLOW Calf Cuff, w/Foam (White) 19" Max. Calf Circumference Ref. 3010-PL Lot 122692, 122715

FDA Recall Z-1638-2013 · Posted 2013-07-01

Recall StatusTerminated
Product CodeJOW
Date Initiated2013-05-03
Date Terminated2013-07-11
Product Quantity229

Reason for Recall

The seal which maintains a sterile barrier for reprocessed medical devices became compromised to the point where product may fall out of the pouch.

Firm Action

MEDISISS sent a Correction Immediate Action Request letter dated May 2, 2013, to all affected customes. The letter identified the product, the problem, and the action to be taken by the customer. The letter directed the customer to inspect the products prior to use for seal integrity as specified on the product labeling: "Sterile unless package has been damaged or opened. Do not use if package is damaged." This will be followed up by a visit from the customer's sales representatives who are tasked with inspecting all the lots that may be remaining in the customer's inventory. The sales representative is directed to complete a "Verification of Seal Inspection" form and return it to MEDISISS Customer Relations. Devices with unacceptable seals are to be returned to MEDISISS. The sales representative is responsible for obtaining a shipping label from customer service to enable the devices to be returned to the recalling firm via Fed Ex. Customers with questions were instructed to call 866-866-7477.

Root Cause

Packaging process control

Distribution Pattern

Nationwide Distribution.