Restoration Modular Cone Bodies; titanium Alloy and Hydroxyapatite, Sterile. Catalog number 6276-1-021, Catalog number 6276-1-025, Catalog number 6276-1-219, and Catalog number 6276-1-325. Howmedica Osteonics Corp, Stryker Ireland, Carrighwohill Sterile. Intended Use: For primary or revision total hip arthroplasty and severe proximal bone loss. Stems are intended to be used with Howmedica Osteonics femoral heads, unipolar and bipolar components, and acetabular components. Conditions for which product may be appropriate for use: rheumatoid arthritis, correction of deformity, non-inflammatory degenerative joint disease, revision surgery, treatment of non-union, fracture of proximal femur with head involvement.

FDA Recall Z-1642-2009 · Posted 2009-05-21

Recall StatusTerminated
Product CodeLZO
Date Initiated2009-04-02
Date Terminated2012-08-27
Product Quantity17 units

Reason for Recall

Packaging burst test values for the Restoration Modular Cone Body family were found to be incorrectly transferred from the finalized validation to the work specification. Testing showed that burst test results for recalled products fell below the validated parameters.

Firm Action

Urgent Product Recall letters were sent via Federal Express on April 2, 2009 to Stryker branches, Chief of Orthopaedics, Hospital Rick Management and Surgeons. The letter stated the issue, the potential hazards, and the risk mitigation. Customers were asked to examine their inventory and hospital locations to identify the affected products; retrieve all effected products and return it to their branch or agency warehouse for reconciliation; reconcile all products on the attached Product Accountability Form and fax a copy to the number listed; return the affected product via the Inter-Org Transfer and ship it to the attention of the individual listed; attach the fluorescent orange PRODUCT REMEDIATION sticker to their return, indicating the Product Remediation #, and mark the box with the words "PRODUCT RECALL". Questions should be directed to Colleen O'Meara at 201-831-5970.

Root Cause

Process control

Distribution Pattern

Nationwide Distribution