Terminated

- Turbosonics Mini-Flared ABS Tip, 45 Degree OZil 12 Degree, (Purplish/ Pink) 0.9mm Ref. 8065751177 Lot 123215 - TurboSonics Mini-Flared ABS Tip, 20 Degree Tip Angle, 45 Kelman, (Purplish Pink/Silver) 0.9mm Ref. 8065750853 Lot 122590, 123372 - TurboSonics Tapered ABS MicroTip" 30 Kelman, (Purple/Silver) 0.9mm Ref. 8065750263 Lot 122309 Series 20000" Legacy TurboSonics Flared ABS MicroTip, 30 Degree, Kelman, (Pink) 0.9mm Ref. 8065740839 Lot. 122309 Phacofragmentation system.

FDA Recall Z-1642-2013 · Posted 2013-07-01

Recall StatusTerminated
Product CodeNKX
Date Initiated2013-05-03
Date Terminated2013-07-11
Product Quantity90

Reason for Recall

The seal which maintains a sterile barrier for reprocessed medical devices became compromised to the point where product may fall out of the pouch.

Firm Action

MEDISISS sent a Correction Immediate Action Request letter dated May 2, 2013, to all affected customes. The letter identified the product, the problem, and the action to be taken by the customer. The letter directed the customer to inspect the products prior to use for seal integrity as specified on the product labeling: "Sterile unless package has been damaged or opened. Do not use if package is damaged." This will be followed up by a visit from the customer's sales representatives who are tasked with inspecting all the lots that may be remaining in the customer's inventory. The sales representative is directed to complete a "Verification of Seal Inspection" form and return it to MEDISISS Customer Relations. Devices with unacceptable seals are to be returned to MEDISISS. The sales representative is responsible for obtaining a shipping label from customer service to enable the devices to be returned to the recalling firm via Fed Ex. Customers with questions were instructed to call 866-866-7477.

Root Cause

Packaging process control

Distribution Pattern

Nationwide Distribution.