Amplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080
FDA Recall Z-1644-2025 · Posted 2025-04-21
Advertisement
Recalling FirmAbbott
Recall StatusOpen, Classified
Product CodeDQY
Date Initiated2025-03-17
Product Quantity2683
Reason for Recall
Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism.
Firm Action
On March 17, 2025, the firm notified customers via hand-delivered customer communications titled "Urgent Medical Device Recall."
Customers were instructed to share the notice with applicable personnel and to return any remaining unused devices from impacted lots. Abbott representatives will assist with product return and replenishment.
If you have any questions, please contact your local Sales Representative or Abbott Support at 1-800-544-1664.
Root Cause
Process control
Distribution Pattern
Worldwide - US Nationwide distribution.