Terminated

7F Prelude Short Sheath Introducer, 4CM in Length with Marker Tip and 0.038" X 50CM Guide Wire; REF No.: PSS-7F-4-038MT, Sterile EO, Merit Medical Systems, Inc., South Jordan, Utah 84095. To provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries while maintaining hemostasis for a variety of diagnostic and therapeutic procedures. This device can also provide access to a native or synthetic graft used for hemodialysis. The side port of the sheath allows adequate flow to perform temporary hemodialysis.

FDA Recall Z-1647-2009 · Posted 2009-06-10

Recall StatusTerminated
Product CodeDYB
Date Initiated2009-04-14
Date Terminated2010-02-09
Product Quantity1072 units

Reason for Recall

External tubing may separate during use, allowing blood loss.

Firm Action

All of Merit's US sales representatives and distributors were notified by Prelude Short Sheath Introducer - Product Recall letter on 4/14/2009. They were instructed to promptly contact their customers. Customers were instructed to isolate, remove from inventory, and return all affected inventory. A Product Retrieval Form was to be returned to Merit. Further information is available at 801-316-4998 or 801-208-4344.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution -- USA, including states of CA, CO, GA, IL, MO, OK, and OR and countries of Austria, Belgium, Canada, Denmark, France, Germany, Hong Kong, Japan, Latvia, Lithuania, Netherlands, New Zealand, Spain, Sweden, and UK.