Covidien Kangaroo Connect Enteral Feeding Pump Item Number: 384400 (US) Intended to provide enteral nutrition to a patient for hospital and acute care settings.
FDA Recall Z-1648-2016 · Posted None
Advertisement
Recalling FirmMedtronic
Recall StatusTerminated
Product CodeLZH
Date Initiated2016-04-01
Date Terminated2018-05-18
Product Quantity223 units
Reason for Recall
Kangaroo Connect Feeding Pump Occlusion alarms fail to alarm
Firm Action
Medtronic(formerly) Covidien notified accounts via Urgent Medical Device Recall letter, dated 4/1/16, that specific serial numbers with software versions 1.12 and 1.13 have a software anomaly. The firm shipped replacement units with software V 1.14. Affected devices must be returned to the firm once the replacement units have been received. A Recalled Product Return Form required to be completed and returned. Questions or concerns, contact Medtronic representative or Customer Service at (800) 882-5878.
Root Cause
Device Design
Distribution Pattern
Worldwide Distribution -- USA, Australia, Canada, France, and Singapore.