Open, Classified

Puracol and Puracol Plus Collagen Wound Dressings: MSC8522 DRESSING,COLLAGEN,PURACOL,STRL,2" X 2" MSC8622EP DRESSING,COLLAGEN,PURACOL PLUS, 2X2.2"

FDA Recall Z-1650-2026 · Posted 2026-03-27

Recall StatusOpen, Classified
Product CodeKGN
Date Initiated2026-02-10
Product Quantity106,621 US; 2,350 OUS

Reason for Recall

Elevated endotoxin levels identified in affected lots. Issue may elicit strong inflammatory response if used, including local tissue irritation, delayed wound healing, or systemic effects such as fever and hypotension. In severe cases, septic shock may result.

Firm Action

Firm began notifying customers on 2/10/26 via "Urgent Medical Device Recall" letters. Customers were instructed to immediately locate and quarantine all affected product. Affected product must be destroyed. Customer should respond to the firm to confirm quantity of affected product in inventory. The customer will receive credit from the firm once the response is submitted. Distributors should notify any of their affected downstream accounts. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Root Cause

Under Investigation by firm

Distribution Pattern

Domestic distribution nationwide US. International distribution to Mexico and Panama.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →