Hitachi CXR4 Computed Tomography (CT) Scanners, Software Version 1.51 or less.
FDA Recall Z-1651-2009 · Posted 2009-07-07
Advertisement
Recalling FirmHitachi Medical Systems America Inc
Recall StatusTerminated
Product CodeJAK
Date Initiated2009-04-14
Date Terminated2012-09-21
Product Quantity47 units
Reason for Recall
Incorrect scale on image: A software error can occur if two (2) different Field-of-View (FOV) settings are registered in a protocol using Multi-Reconstruction. This could cause the image generated to be displayed with the wrong scale factor rendering anatomical measurements incorrect.
Firm Action
HMSA issued an URGENT - DEVICE CORRECTION notification to their customers on 4/14/2009. The notification describes the potential software error and the condition under which these image distortions can occur. The notification goes on to provide the customer with a work-around procedure which they can use to avoid encountering this issue with their current software configuration(s). The letter states that the firm will be installing updated software on all affected systems beginning in June, 2009, and that a Hitachi Service Rep. will be contacting each customer and making arrangements to come out and install a software upgrade on the affected system(s) at no cost to the customer. Lastly, the letter provides a telephone # to the firm's Applications Helpline: 1-877-228-2777, which the customer can use to obtain additional information, if desired.
Root Cause
Software design
Distribution Pattern
Nationwide distribution, including Puerto Rico (PR).
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement