Lithium Thionyl non-rechargeable disposable battery pack P/N 350-0294-01 used with ER920 Wireless Series Arrhythmia Event Recorder. Non-rechargeable disposable battery pack.
FDA Recall Z-1652-2013 · Posted 2013-07-02
Advertisement
Recalling FirmBRAEMAR, INC.
Recall StatusTerminated
Product CodeDRG
Date Initiated2012-02-24
Date Terminated2013-07-15
Product Quantity3286 packs
Reason for Recall
Braemar has recently became aware of a battery related incident that occurred with EWT battery packs that are used in a Braemer ER920W wireless event monitor and Fusion device.
The event did not result in any patient related injury or present any clinical impact.
Firm Action
Braemar began notifying customers by telephone on 2/24/2012. Braemar issued Advisory Notice letter P/N 600-066201 RevA on 3/02/2012. The letter described the product, the problem, and the action to be taken by the customer. Customers were requested to only use TESTED batteries that they have receivewd fromBraemar starting on February 23, 2012. Complete and return the attached Verification Form. If customers have any untested product in their inventory, they were instructed to segregate those battery packs and await further instruction from Braemar to coordinate the return of those untested units. Upon receipt of the attached verification form, Braemar will continue to send replacement battery packs that have passed the additional testing protocol. Customers were instructed to contact their Braemar sales representative if they have any questions regarding the notice.
For questions regarding this recall call 651-286-8620.
Root Cause
Material/Component Contamination
Distribution Pattern
Nationwide Distribution including FL, IL, LMD, MA, MN, NY, OH, PA, TN, and TX.