Terminated

Siemens Linear Accelerator (LINAC) models. Product Usage: deliver X-ray radiation for therapeutic treatment of cancer

FDA Recall Z-1652-2014 · Posted 2014-05-23

Recall StatusTerminated
Product CodeIYE
Date Initiated2014-04-01
Date Terminated2015-03-16
Product Quantity9

Reason for Recall

Siemens Radiation Oncology became aware that customers may be using the Siemens Linear Accelerator in combination with stereotactic accessories which have not been validated as being compatible with Siemens LINAC models.

Firm Action

Siemen sent an Urgent Medical Device Correction letter to all affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local Elekta representative.

Root Cause

Labeling design

Distribution Pattern

US Nationwide Distribution in the states of AL, ME, PA, IL, and CA.