Factor V Leiden kit catalog number 03610179001, Roche Diagnostics, Indianapolis, IN The Factor V Leiden Kit allows the detection and genotyping of a single point mutation (G to A at position 1691) of the human Factor V gene, referred to as Factor V Leiden mutation, from DNA isolated from human whole peripheral blood. The test is performed on the LightCycler 2.0 Instrument utilizing polymerase chain reaction (PCR) for the amplification of Factor V DNA recovered from clinical specimens and fluorogenic target-specific hybridization for the detection of genotyping of the amplified Factor V DNA. The Factor V Leiden test is an in vitro diagnostic test for the detection and genotyping of the Factor V Leiden mutation as an aid to diagnosis in the evaluation of patients with suspected thrombophilia. The test is intended to be used on the LightCycler 2.0 Instrument using the LightCycler Software 4.05 or 4.1.
FDA Recall Z-1653-2011 · Posted 2011-03-14
Advertisement
Reason for Recall
Firm Action
Root Cause
Software design
Distribution Pattern
Worldwide distribution: AL, AZ, CA, CO, DC, FL, GA, HI, IL, KY, LA, MA, MD, ME, MN, MO, MT, NC, ND NE, NH, NJ, NM, NY, OH, OK, OR, SC, SD, TN, TX, UT, VA, VT, WI and Puerto Rico; and country of Europe.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.