ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.

FDA Recall Z-1653-2026 · Posted 2026-04-07

Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2026-03-10
Product QuantityU.S. 758, OUS 2077

Reason for Recall

During patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient. Despite the reported dose showing significantly higher, the actual applied x-ray dose levels are in accordance with the system settings and imaging conditions, as well as the applicable regulatory requirements.

Firm Action

Siemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice, dated March 9, 2026, to all affected customers.

Root Cause

Radiation Control for Health and Safety Act

Distribution Pattern

Worldwide - US Nationwide distribution.