Philips Allura Xper Systems with Velara Generator, x-ray system. FCO72200184.

FDA Recall Z-1654-2011 · Posted 2011-03-16

Recall StatusTerminated
Product CodeIZI
Date Initiated2010-10-07
Date Terminated2012-02-23
Product Quantity1556 units distributed in US (1452-3107=1556)

Reason for Recall

A wire in the connector of the Generator Grid Switch Supervisor box might short-circuit between 12V power and ground, and it will cause a System breakdown.

Firm Action

On 10/15/2010, Philips began sending out the URGENT-Device Correction Notice Failure Causing the Generator to Stop Working notification letter to their consignees. The consignees were informed that a short circuit may occur due to a faulty wire connection in the Supervisor Box Connector in the Generator, and it will cause the generator stops working. When the system fails during a critical interventional case, this may cause risk to patients (no Fluoro available). The systems involved in this recall are Allura Xper with the Velara Generator, Allura Integris with OMCP Generator, MultiDiagnost Eleva, and OmniDiagnost Eleva Imaging system. Consignees can contact Philips Healthcare Call Center at 800-722-9377, #5 and Reference FCO 70800114, 70800115, 72200184, and 72200185.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Devices were distributed through out the US.