Elekta XiO Radiation Treatment Planning System

FDA Recall Z-1655-2010 · Posted 2010-05-26

Recall StatusTerminated
Product CodeMUJ
Date Initiated2010-04-15
Date Terminated2012-06-07
Product QuantityThe release is being used by 34 customers that also have an Elekta linear accelerator

Reason for Recall

Under certain circumstances, XiO will over-estimate dose in the region where the backup jaw is inside the MLC opening

Firm Action

Elekta issued an undated "Important Safety Notice" to all affected sites via e-mail on 4/20/2010. The notice was also sent to the sites on 4/6/10 and 4/12/10 (for 4 sites) before the firm made the decision to recall. The letter explains the correct functioning of the system, the problem, when and why it occurs, the clinical impact, and the work around. Although not stated in the Safety Notice, a self-addressed postcard was enclosed for the customer to return as evidence they received and understood the notice. Sites that do not return the card will be contacted by phone.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution: USA, Canada, Australia, Spain, Italy, the Netherlands, United Kingdom, Taiwan, Malaysia, France, Ireland, and Israel.