ArthroCare brand Topaz XL ArthroWand with Integrated Cable Wand, Catalog Number: AC4045-01; Product is manufactured and distributed by ArthroCare Corporation, Sunnyvale, CA - Sterile Surgical Device. Product is indicated for ablation and debridement of tendons in orthopedic procedures.

FDA Recall Z-1657-2008 · Posted 2008-09-16

Recall StatusTerminated
Product CodeHRX
Date Initiated2008-01-31
Date Terminated2011-01-08
Product Quantity2,129

Reason for Recall

Devices for which sterility may be compromised as evidenced by punctures in the plastic tray during shipping. The risk to the patient is a potential infection subsequent to the surgical procedure.

Firm Action

Urgent Product Recall letters (Dated January 31, 2008) were issued to end users of the device (medical professionals) and direct distributors. The letter instructs users to return any affected products.

Root Cause

Packaging process control

Distribution Pattern

Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, ME, MN, MO, NC, ND, NJ, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI, and countries of Sweden, Australia, Germany, Italy, UK, Austria, Saudi Arabia, Belgium, Portugal, UAE, Kuwait, Singapore, Ireland, Spain, Finland, South Korea, Russia and Vietnam.