Terminated

Coulter LH 500 Series System. Beckman Coulter, Inc. Brea, CA. Quantitative, automated hematology analyzer and leukocyte differential cell counter for in vitro diagnostic use in clinical laboratories.

FDA Recall Z-1657-2010 · Posted 2010-05-26

Recalling FirmBeckman Coulter Inc.
Recall StatusTerminated
Product CodeGKZ
Date Initiated2010-02-04
Date Terminated2014-09-02
Product Quantity1520 units

Reason for Recall

Issue 1: The system will substitute or omit the characters # @ [\] ' {I) when the system is configured in a language other than English or Chinese. Example: Using the character @, a substitution in the string '123@456' would result in '123#456' and an omission would result in '123456' Issue 2: The system will omit the character ~ regardless of language. Impact: There is a potential for Specimen

Firm Action

Beckman Coulter, Inc. issued "Urgent: Product Corrective Action" notifications dated February 12, 2010. Consignees were informed of the affected product and instructed on proper actions. The firm also asked that the enclosed response forms be returned. For further information, contact your local Beckman Coulter Representative or call 1-800-526-7694.

Root Cause

Software design

Distribution Pattern

Distributed Nationwide and Canada.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →