Terminated

***REF SC2101***eMax 2 Plus Console, Single Port, with Irrigation***SN G17307901802*** Product Usage: The RPM of the Motor Handpiece may be less than 80,000 RPM (approximately 72,000 RPM) in the free running state when using the optional Hand Control. The low Motor Handpiece RPM would only be experienced in the Free running state (Hand Control fully depressed and no load on the cutting bur). This condition would not affect the power output of the Handpiece in the loaded working range (typically 50,000 to 70,000 RPM.

FDA Recall Z-1658-2013 · Posted 2013-07-02

Recall StatusTerminated
Product CodeNAV
Date Initiated2011-06-24
Date Terminated2013-12-24
Product Quantity61 units

Reason for Recall

The RPM of the Motor Handpiece may be less than 80,000 RPM.

Firm Action

Anspach Effort sent an "URGENT: Medical Device Correction" letter dated June 24, 2011 to their customers via mail. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete a reply form promptly confirming receipt of their letter and return it by fax to the number provided on the form. If any of the affected products were distributed to other services or facilities, please forward this information as appropriate. For questions contact Anspach Customer Support at 800-327-6887 or email [email protected].

Root Cause

Process control

Distribution Pattern

Worldwide Distribution - USA Nationwide including the states of: AZ, CA, FL, KY, ME, MI, MO, OH, PA, TN, TX , VA, and the countries of: Australia, Egypt, Italy, Japan, Switzerland, Taiwan, and Turkey;