Liquid Waste Bottle, Catalog#10291587. The Immulite 2000 systems are continuous random-access instruments that perform chemiluminescent immunoassays. These instruments use serum, plasma, or urine samples for in vitro diagnostic testing.
FDA Recall Z-1659-2015 · Posted 2015-05-23
Advertisement
Recalling FirmSiemens Medical Solutions Diagnostics
Recall StatusTerminated
Product CodeJJE
Date Initiated2015-03-05
Date Terminated2018-01-16
Product Quantity231 bottles
Reason for Recall
The bottle assemblies were manufactured with a quality issue where in some cases the smaller diameter opening on both IMMULITE 2000 and IMMULITE 2000XPi Water Bottles and Liquid Waste Bottles is undersized and/or deformed, preventing the cap from closing or fastening securely to the bottle.
Firm Action
Siemens Healthcare Diagnostics sent an Urgent Device Correction Letter/Field Correction Effective Check Form dated 3/5/15 via FedEx to their affected customers.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide Distribution.