GE Healthcare, Innova 3100 IQ : GE Innova 3100 / 3100 IQ, Cardiovascular Imaging System, (The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector)

FDA Recall Z-1660-2010 · Posted 2010-05-26

Recalling FirmGe Healthcare, Llc
Recall StatusTerminated
Product CodeMQB
Date Initiated2010-03-24
Date Terminated2013-05-21
Product Quantity75

Reason for Recall

A potential software failure related to the Worklist functionality has been identified on a subset of Innova 2100IQ, 3100IQ and 4100IQ systems. When using the Worklist function to import patient data to Innova, there are two scenarios that will cause the next exam to fail and the inability to recall acquired sequences.

Firm Action

Consignees were sent on 3/24/10, a GE Healthcare " Urgent Medical Device Correction" letter dated March 23, 2010. The letter included the Safety Issue, Affected Product Details, Product Correction and Contact Information.

Root Cause

Software design

Distribution Pattern

Worldwide distribution: USA, Puerto Rico, VENEZUELA, UNITED KINGDOM, TURKEY, THAILAND, TAIWAN, PROVINCE OF CHINA, SWEDEN, SPAIN, RUSSIAN FEDERATION, ROMANIA, POLAND, PHILIPPINES, NORWAY, NETHERLANDS, MOROCCO, REPUBLIC OF KOREA, KAZAKSTAN, JAPAN, ITALY, ISLAMIC REPUBLIC OF IRAN, INDONESIA, INDIA, GERMANY, FRANCE, FINLAND, EGYPT, CHINA, CHILE, CANADA, BRAZIL, BELGIUM, BELARUS, AUSTRALIA, and ARGENTINA.