Beckman Coulter DxI 9000 Access Immunoassay Analyzer

FDA Recall Z-1664-2025 · Posted 2025-04-24

Recalling FirmBeckman Coulter, Inc.
Recall StatusOpen, Classified
Product CodeJJE
Date Initiated2025-03-27
Product Quantity370 units

Reason for Recall

When DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose communication with the host system. The lost connection interrupts sample processing, which delays reporting patient test results and may subsequently delay patient treatment.

Firm Action

Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 03/27/2025 via USPS first class mail. The notice explained the issue with the device, impact to the patient, and provided a workaround until a software upgrade is available. If the consignee further distributed the affected device, they were directed to provide a copy of the notice to the recipient. For questions, contact Customer Support Center, from the website: http://www.beckmancoulter.com

Root Cause

Software design

Distribution Pattern

Worldwide distribution.