SIEMENS Axiom Luminos dRF Max, a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image.
FDA Recall Z-1668-2015 · Posted 2015-05-23
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeOWB
Date Initiated2015-04-13
Date Terminated2017-03-20
Product Quantity133 total
Reason for Recall
Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.
Firm Action
A safety advisory notice, dated April 13, 2015, was sent to end users with affected devices to alert them about the issue and associated risks,and to provide instructions for avoiding potential risk. A new software version will be released to permanently resolve the issue.
Root Cause
Software design
Distribution Pattern
Nationwide Distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.