Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

CADD-Solis VIP Ambulatory Infusion Pump; Model/Number: 21-2120-0100-02 21-2120-0100-03 21-2120-0100-06 21-2120-0100-232 21-2120-0100-233 21-2120-0100-50 21-2120-0100-51 21-2120-0100-95 21-2120-0102-02 21-2120-0102-03 21-2120-0102-06 21-2120-0102-07 21-2120-0102-08 21-2120-0102-12 21-2120-0102-13 21-2120-0102-14 21-2120-0102-15 21-2120-0102-17 21-2120-0102-237 21-2120-0102-238 21-2120-0102-247 21-2120-0102-249 21-2120-0102-51 21-2120-0102-78 21-2120-0102-92 21-2120-0102-97 21-2120-0102-98 21-2120-0103-01 21-2120-0103-02 21-2120-0103-03 21-2120-0103-06 21-2120-0103-07 21-2120-0103-08 21-2120-0103-12 21-2120-0103-14 21-2120-0103-15 21-2120-0103-17 21-2120-0103-249 21-2120-0103-51 21-2120-0103-78 21-2120-0104-01 21-2120-0104-231 21-2120-0105-00 21-2120-0105-01 21-2120-0105-02L 21-2120-0105-03L 21-2120-0105-06L 21-2120-0105-07L 21-2120-0105-08 21-2120-0105-08L 21-2120-0105-12L 21-2120-0105-13L 21-2120-0105-14L 21-2120-0105-15L 21-2120-0105-17L 21-2120-0105-50 21-2120-0105-50L 21-2125-0104-01 21-2125-0105-03L 21-2125-0105-07L 21-2125-0105-08L 21-2125-0105-50L 21-2127-0104-01 21-2127-0105-01 21-2127-0105-02L 21-2127-0105-03L 21-2127-0105-06L 21-2127-0105-08L 21-2127-0105-50L

FDA Recall Z-1668-2025 · Posted 2025-05-07

Recall StatusOpen, Classified
Product CodeFRN
Date Initiated2025-04-10
Product Quantity185,975 US; 127,701 OUS Total

Reason for Recall

Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.

Firm Action

On April 10, 2025, the firm notified affected consignees via letters titled "Urgent Medical Device Correction." Customers were informed of conditions under which the false Upstream Occlusion Alarm may occur. The USO alarm can be cleared during these conditions by removing the administration from the pump, re-attaching the administration set, and restarting delivery. Customers should verify there is no upstream occlusion present. The firm will be making an update to the Infusion Pump Operator's Manual to include these conditions.

Root Cause

Labeling design

Distribution Pattern

Worldwide - US nationwide distribution.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑