Terminated

Cardio Medical Products, Inc. Defibrillation/Cardioversion/Monophasic or Biphasic Pacing/ECG Electrode Rx Only Single Patient Use Only Manufactured for Cardio Medical Products 385 Franklin Avenue, Suite L Rockaway, New Jersey 07866 www.cardiomedicalproducts.com (800) 227-3633 (973) 586-9624 FAX Defibrillator

FDA Recall Z-1669-2013 · Posted 2013-07-08

Recall StatusTerminated
Product CodeLDD
Date Initiated2013-04-26
Date Terminated2016-08-15
Product Quantity657 units

Reason for Recall

Covidien received customer reports of Arcing/Sparking on the defibrillation electrode leadwire which may affect Cardio Medical Products Inc. defibrillator electrodes Model Number O101.

Firm Action

Cardio Medical Products, Inc. issued an Urgent Medical Device Recall letter dated April 8, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review their inventory and segregate any product with the affected lot numbers and return affected product according to the instructions for product credit by contacting the Customer Service group at 1-800-962-9888, option 1, then option 2, and reference the recall letter to request a Return Goods Authorizing (RGA). Customers were instructed to complete the enclosed Verification Form and return it to Covidien by fax to 800-895-6140, even if they did not have any of the affected lots. Customers were also instructed to report any issues to the Quality Assurance Group to ensure proper device reporting procedures are followed. Customers with questions were instructed to call 1-800-962-9888, option 8, then ext 2500. For questions regarding this recall call 973-586-9500.

Root Cause

Process control

Distribution Pattern

Nationwide Distribution