The FastPack System Comlete Immunoassay Kit Immunoassay for the Quantitative Measurement of SHBG is a chemiluminescent immunoassay intended for the quantitative determination of Sex Hormone Binding Globulin in human serum and plasma on the FastPack System. The FastPack IP SHBG assay is intended for use as an aid in the diagnosis of androgen disorders. Each Kit Contains 30 FastPack IP SHBG Immunoassays 1 Calibrator 1 x 3 mL Vial 1 Control 1 1 x 3 mL Vial 1 Control 2 1 x 3 mL Vials 32 Sample Diluent 0.9 mL Vials REF 25000081 IVD CE 1908038-1P 2020-02-22 (01) 20816467020345 (17) 200222 (10) 1908038-1P

FDA Recall Z-1672-2020 · Posted None

Recalling FirmQualigen Inc
Recall StatusTerminated
Product CodeCDZ
Date Initiated2019-11-06
Date Terminated2020-09-23
Product Quantity24 Kits (30 test packs each) 720 tests

Reason for Recall

The chemiluminescent immunoassay has been noted to have higher than expected results which may result in inaccurate quantitative determinations of Sex Hormone Binding Globulin in human serum and plasma.

Firm Action

On October 31, 2019, a "Urgent: Product Recall" letter was sent to customer via E-Mail. In addition to informing consignees about the recall the customer notification asked consignees to take the following actions: 1. The firm will provide replacement kits at no cost as soon as they become available. 2. For any kits that have been purchased a refund can be requested in place of replacement kits. 3. As part of the product replacement process, please confirm that any remaining SHBG FastPacks/Kits of Lot Number 1908038-1P have been discarded on the form below. 4. The firm will call to confirm that you have received the letter and provide any additional details you may require. For questions, please contact System Support at [email protected].

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide distribution in the states of AL, AK, CA, HI, and TX.