Mallory Head Calcar Femoral 55 mm resection/porous coated 15.0 mm x 220 mm left stem, Sterile, REF 11-104972, Biomet Orthopaedics, Warsaw, IN. Implant device intended for use in femoral revision surgery.
FDA Recall Z-1675-2010 · Posted 2010-05-26
Recalling FirmBiomet, Inc.
Recall StatusTerminated
Product CodeLPH
Date Initiated2010-03-10
Date Terminated2010-10-18
Product Quantity6
Reason for Recall
The firm has found that some lots of Mallory Head Revision Stems had flute geometry that was out of specification. The flute geometry was too deep.
Firm Action
Biomet Orthopedics issued an "Urgent Medical Device Recall Notice" dated March 10, 2010. The letter went to distributors and users of the recalled device. The letter instructed the consignees to locate the recalled product and return the product to the firm.
For further information, contact Biomet at 1-800-348-9500 or 1-574-372-3983, Monday through Friday, 8:00 AM to 5:00 PM.
Root Cause
Process change control
Distribution Pattern
Worldwide Distribution -- United States (OH, IN, FL, KY, NV and CA), Mexico, Chile, China, Japan and The Netherlands.