Terminated

Toshiba DUA-450F Kalare Radiographic Tilting Table.

FDA Recall Z-1675-2011 · Posted 2011-03-16

Recall StatusTerminated
Product CodeIXR
Date Initiated2010-07-15
Date Terminated2012-04-20
Product Quantity238 systems

Reason for Recall

Top table lateral movement on their DUA-450F Kalare Table cannot be performed, or the table top lateral movement is performed at an extremely low speed. As a result, positioning of the patient cannot be performed by tabletop movement.

Firm Action

Toshiba America Medical Systems (TAMS) issued Urgent Medical Device Correction notification dated July 8, 2010, to consignees via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS as proof of notification.

Root Cause

Software design

Distribution Pattern

US Nationwide