Wound Drain - Label reads in part "Elutia ***Hubless Coated Silicone Surgical Wound Drain***" Two different sizes, 10 units per box: 19 Fr Round Hubless part number 7110; 24 Fr Round Hubless part number 7118. AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system.
FDA Recall Z-1680-2014 · Posted 2014-06-02
Recalling FirmBacterin International, Inc.
Recall StatusTerminated
Product CodeOEI
Date Initiated2014-05-15
Date Terminated2014-06-02
Product Quantity390
Reason for Recall
The recalling firm reported there was inadvertent omission of quarterly (Q4 - 2013 and Q1 - 2014) dose audits/verifications on the affected products. All of the affected products were appropriately sterilized
Firm Action
Bacterin sent e-mail and FedEx letter with return receipt request to it's single consignee on May 15, 2014. Consignee is asked to separate the recalled product in a secure location for return to Bacterin International, Inc. Consignee is asked to complete tracking/verification form included in the letter/e-mail and call 1-888-886-9354 from 8:00AM to 5:00PM (Mountain Time) or send an e-mail to [email protected] to obtain a Return Authorization (RA) number.
Root Cause
Use error
Distribution Pattern
Distributed to one foreign consignee in Lebanon.