Terminated

Altrus Thermal Tissue Fusion, Tissue Fusion Handpiece 5 mm O.D., 16 cm Shaft, Catalog Number 60-9520-001, Sterile/EO. The Altrus Thermal Tissue Fusion System is comprised of a dedicated energy source and disposable handpiece used to ligate (seal) and divide (cut) blood vessels and tissue bundles that fit into the jaws of the handpiece.

FDA Recall Z-1681-2012 · Posted 2012-05-29

Recalling FirmConMed Electrosurgery
Recall StatusTerminated
Product CodeGEI
Date Initiated2012-05-07
Date Terminated2012-06-05
Product Quantity18 units

Reason for Recall

The devices were distributed prior to the firm completing performance qualifications for their package sealing and adhesive bonding.

Firm Action

ConMed Electrosurgery sent an "URGENT: DEVICE RECALL" letter dated May 2, 2012 to all affected customers. The letter identifies the product, problem,and actions to be taken by the customers. The letter instructs customers to review their inventory and to complete and return a Business Reply Form via fax to 303-699-9854.

Root Cause

Process control

Distribution Pattern

Class II Recall-Nationwide Distribution-including the states of OH, GA, IL, PA, AZ, KY, FL, NY, MO, TN, TX, AL, IN, SC, MD, CT, NJ, and CA.