Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO AR2 (5PK), REF H74908526862; cardiac catheter
FDA Recall Z-1681-2024 · Posted 2024-05-10
Advertisement
Recalling FirmBoston Scientific Corporation
Recall StatusOpen, Classified
Product CodeDQO
Date Initiated2024-03-21
Product Quantity9271 catheters
Reason for Recall
An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delaminating and in some instances material detachment in the catheters inner lining.
Firm Action
Boston Scientific issued an Urgent Medical Device Removal notice to its consignees on 03/21/2024 via overnight delivery. The notice explained the issue, the potential risk, and requested the removal of the affected products. Distributors were directed to notify their consignees.
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide distribution. US, US Territories, EMEA, Canada, LATAM and Asia Pacific countries/regions.