Ingenuity TF Positron Emission Tomography (PET) and Magnetic Resonance Imaging (MRI), 882380, Philips Medical Systems. The device is a diagnostic imaging system for fixed installations that combines Positron Emission Tomography (PET) and Magnetic Resonance Imaging (MRI).

FDA Recall Z-1682-2015 · Posted 2015-05-28

Recall StatusTerminated
Product CodeKPS
Date Initiated2015-03-11
Date Terminated2018-06-26
Product Quantity10 units

Reason for Recall

It was noticed that the spectroscopy voxel map did not align with the anatomical images of the phantom. This resulted in incorrect orientation of the Chemical Shift Imaging (CSI) data.

Firm Action

On 3/11/2015 the firm sent Urgent - Medical Device Correction letters to their customers. The letters identified the affected device, explained the reason for the recall, how to identify affected product, advice on actions to be taken by customer/user, and any actions planned by the firm. Customers were advised to use software provided by Philips to view spectroscopy images; and, not to export and view images using third party software. The firm stated that it will be issuing a software update. If further information or support concerning the issue is needed, customers are asked to contact their local Philips representative or local Philips Healthcare office.

Root Cause

Software in the Use Environment

Distribution Pattern

Worldwide Distribution -- US, including the states of NY, OH, & TX; and, the countries of Finland, Germany, Japan, Netherlands, South Korea, Spain, & Switzerland.