RUSCH LASERTUBE (Rubber), Order No. : 102004-000040, REF 102004, SIZE I.D. mm 4,0
FDA Recall Z-1682-2019 · Posted None
Advertisement
Recalling FirmTeleflex Medical
Recall StatusTerminated
Product CodeBTR
Date Initiated2019-04-01
Date Terminated2025-06-03
Product Quantity1787 units
Reason for Recall
The firm received reports indicating that the laser guard foil partially separated and/or slightly detached at the edges.
Firm Action
The firm initiated the recall by letter on 04/01/2019. The letter stated reason for the recall and requested the return of the recalled units. Distributors were directed to collect the recalled units from their customers, consolidate them, and complete and return the acknowledgement form. The recalled units will be returned.
Root Cause
Process control
Distribution Pattern
Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.