Open, Classified

McKesson Cardiology ECG Management, Change Healthcare Cardiology ECG Management

FDA Recall Z-1682-2020 · Posted None

Recall StatusOpen, Classified
Product CodeDQK
Date Initiated2020-02-10
Product Quantity51 units

Reason for Recall

Inaccurate estimation of the heart rate may occur in instances where the user is estimating heart rate counting grid squares using the Sequential ECG format

Firm Action

Change Healthcare Canada Company notified customers on about 02/10/2020 via "Urgent Field Safety Notice" letter. Customers were informed of an issue whereby users who are estimating heart rate by counting grid squares, may assume that the Sequential ECG format is presented on a scale of 25 mm/sec and thereby may inaccurately estimate the heart rate value. Instructions included that the Simultaneous ECG format can be used rather than the Sequential ECG format, until the software is updated and the Sequential ECG format option can be used while maintaining awareness to the scale bar on the screen. Customers were also instructed to contact their users and train them on how to verify the ECG format mode they are working on and how they can change it. The notice needs to be distributed to all those within their organization who need to be aware and to please alert other parties affiliated with your organization that are affected by this field safety notice. Additionally, customers were informed that Change Healthcare is developing a software update that will enable presentation of the Sequential ECG format view using the 25 mm/sec scale and were requested to contact Change Healthcare Customer Support to: 1) Acknowledge they have read and understood this Urgent Field Safety Notice and 2) Arrange installation of software update/upgrade. Change Healthcare Customer Support contacts: - US / Canada: 1 866-777-0202 - UK / Ireland: +44 (0) 208 9527399

Root Cause

Device Design

Distribution Pattern

Distribution US nationwide, Canada and the United Kingdom