Terminated

Is-anti-Gliadin IgG Enzyme Immunoassay Test Kit packaged in cardboard boxes with individual components secured within a foam insert. Catalog Number # 720-800. - Laboratory In-Vitro Diagnostic Assay intended for the semi-quantitative detection of IgG antibodies to Gliadin in human serum by indirect enzyme immunoassay as an aid in the diagnosis of celiac disease.

FDA Recall Z-1684-2008 · Posted 2008-09-16

Recalling FirmDiamedix Corporation
Recall StatusTerminated
Product CodeMST
Date Initiated2008-03-21
Date Terminated2009-02-11
Product Quantity174 Test Kits

Reason for Recall

Standards and controls in the Is-anti-Gliadin IgG Test Kits are exhibiting low activity. A drop in the O.D.s the Positive Control could potentially recover low out of range (which would invalidate the test), or false positive test results could be obtained.

Firm Action

A Customer Notification Letter was sent to each customer via Certified Mail for domestic customers and via Federal Express for international customer requesting them to contact Diamedix Technical Services to arrange for the return and replacement of any kits of the lot in question. If you have any questions please contact Glenn Gerstenfeld at 1-800-327-4565 or (305) 324-2300.

Root Cause

Environmental control

Distribution Pattern

Class II Recall - Worldwide Distribution.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →