Terminated

Trilogy self-tapping bone screw Bone screw 4.5x25 selftap 00625004525 Bone screw 4.5x50 selftap 00625004550 Bone screw 6.5x15 selftap 00625006515 Bone screw 6.5x20 selftap 00625006520 Bone screw 6.5x25 selftap 00625006525 Bone screw 6.5x30 selftap 00625006530

FDA Recall Z-1684-2016 · Posted 2016-05-16

Recall StatusTerminated
Product CodeLPH
Date Initiated2016-05-04
Date Terminated2017-07-05
Product Quantity3893

Reason for Recall

A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.

Firm Action

All distributors will be notified via electronic mail. Distributors with affected lots will also be notified via courier. Hospitals risk managers and surgeons will be notified via recall notification letter, dated May 4, 2016.

Root Cause

Packaging change control

Distribution Pattern

Worldwide distribution. US nationwide, Korea, Japan, Argentina, Brazil, Singapore, Taiwan, China, El Salvador, Australia, India, Hong Kong, Chile, Germany, Italy, France, Spain, UK, Sweden, Canada.