Integra Hallu-Fix cannulated screw driver (model number P/N 129733ND) and cannulated screwdriver tip (model number P/N 129735ND). Indicated for use in fixation of fractures, osteotomies and arthrodesis of the first metatarso-phalangeal joint.

FDA Recall Z-1685-2013 · Posted 2013-07-11

Recall StatusTerminated
Product CodeLRZ
Date Initiated2013-02-01
Date Terminated2013-07-25
Product Quantity154 screwdrivers were distributed; 67 screwdriver tips were distributed.

Reason for Recall

The firm has received numerous complaints of the cannulated screw driver tip breaking during surgery.

Firm Action

The firm sent recall notification letters to all consignees via traceable email or traceable fax on 02/01/13. The firm requested that the consignees fill out and return the Recall Acknowledgement and Return Form.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution-USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NV, NY, OH, PA, RI, SC, TN, TX, UT, VA, and WA and the countries of Canada, Austria, Belgium, Switzerland, Czech Republic, Germany, Denmark, Spain, Finland, France, Ireland, Israel, Italy, Lebanon, Luxembourg, Netherlands, Norway, Portugal, Russia, Saudi Arabia, Sweden, United Arab Republic, United Kingdom, and South Africa.