The Ultem Polymeric Resin Koh Cup is a non- sterile, reusable vaginal fornices delineator. It is made of medical grade plastics and not made with natural rubber latex. Model Number(s): KCP-30-2
FDA Recall Z-1686-2021 · Posted None
Advertisement
Recalling FirmCooperSurgical, Inc.
Recall StatusOpen, Classified
Product CodeHEW
Date Initiated2021-04-23
Product Quantity6 boxes or 12 units
Reason for Recall
The affected Koh Cup were incorrectly packaged with a 3.5cm Ultem Koh cup, P/N 34131-04, instead of a 3.0cm Ultem Koh cup, P/N 34131-03 as required.
Firm Action
On 04/30/2021, Cooper Surgical issued an Urgent Medical Device Removal notice to customers via letter notifying them Cooper Surgical is issuing this recall due to a labeling discrepancy with respect to the cup size, where a 3.5cm Ultem Cup was placed in the Product rather than a 3.0cm Ultem Cup as indicated by the Part Number on the packaging label.
Root Cause
Labeling design
Distribution Pattern
Domestic: UT, OH, DE
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.