GE Innova 3100-IQ, Cardiovascular Imaging System, GE Healthcare, Waukesha, WI 53188 (The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector). Indicated for use in generating fluoroscopic images of human anatomy for vascular angiography diagnostic and interventional procedures and optionally, rotational imaging procedures. It is also intended for generating fluoroscopic images of human anatomy for cardiology, diagnostic and interventional procedures.

FDA Recall Z-1689-2009 · Posted 2009-09-24

Recalling FirmGe Healthcare
Recall StatusTerminated
Product CodeMQB
Date Initiated2008-10-09
Date Terminated2012-01-07
Product Quantity929 units total

Reason for Recall

GE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration procedure check, GE Healthcare found the AKR calibration procedure does not exactly reflect the worst-case clinical usage of the product. This may result, when using large Equival

Firm Action

Consignees were sent a GE Healthcare "Important electronic Product Radiation Warning" letter dated November 3, 2008. The letter was addressed to Healthcare Administrator/Risk Managers, Managers of Radiology/Cardiology, Radiologists/Cardiologists. The letter described the Defect Description and Related Hazards, Affected Product Details, Safety Instructions, Product Correction and Contact Information. For further information, contact the GE Healthcare Technical Support Center at 1-800-437-1171.

Root Cause

Radiation Control for Health and Safety Act

Distribution Pattern

Nationwide Distribution -- United States including Puerto Rico and Guam.