Brilliance 16 - Model #728246, CT Scanning System is subject to recall/correction Product Usage: The Brilliance 6, 16, 40 64, Big Bore, iCT, iCT SP and Ingenuity CT are Computed Tomography X-Ray Systems that are intended to produce cross-sectional images of the body by computer reconstruction of X-Ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment , patient, and equipment supports, components and accessories.

FDA Recall Z-1690-2012 · Posted 2012-05-30

Recall StatusTerminated
Product CodeJAK
Date Initiated2012-03-26
Date Terminated2013-02-22
Product Quantity10 system units were distributed in the U.S.

Reason for Recall

Component defect. Following their receipt of a customer complaint, Philips Medical determined that the metal blocks which support the scissor slide rails underneath the scanning table (couch) of several of the firm's CT scanning systems were not manufactured to specification and may crack.

Firm Action

Philips issued an 'URGENT Medical Device Correction' letter dated March 27, 2012 to all affected customers. The letter identified the affected products, problem, hazard involved, how to identify affected products, and actions to be taken by customer/user. The letter informed customers that Philips intends to dispatching service personnel to each customer site in order to correct this issue. In the meantime, the letter recommend customer to contact their Philips representative if they hear unusual noises when moving the couch vertically, or if there is unintended vertical motion of the couch. For questions contact your local Philips representative or local Philips Healthcare office. For North America and Canada contact the Customer Care Solutions Center 1-800-722-9377, option 5: Enter Site ID or follow the prompts.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution - U.S. (Nationwide) including the states of: CA, FL,, HI, IA, IN, KY, LA, MA,, MI, MO, MS. NM, NV, OH, PA, PA, SC, TN, TX, WA, and Puerto Rico

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑