1) 24 G x 0.75 in. BD Angiocath" Autoguard" shielded IV catheter (0.7 mm x 19 mm) made of FEP polymer catalog number 381700 2) 24 G x 0.56 in. BD Angiocath-N" Autoguard" shielded IV catheter (0.7 mm x 14 mm) made of FEP polymer catalog number 381720
FDA Recall Z-1690-2016 · Posted 2016-05-18
Advertisement
Recalling FirmBecton Dickinson & Company
Recall StatusTerminated
Product CodeFOZ
Date Initiated2016-03-18
Date Terminated2017-07-19
Product Quantity1,305,150 units
Reason for Recall
The device may have a defect in the catheter. In some instances this defect could result in catheter
separation or breakage.
Firm Action
BD isssued an urgent voluntary product recall lettter/recall response card on 3/18/2016.
Root Cause
Device Design
Distribution Pattern
Nationwide
Official Source