Altius M-INI OCT Posterior Spinal Fixation System, Model Number 1200-9216 The AltiusTM OCT and AltiusTM M-INITM OCT Systems when intended to promote fusion of the cervical spine and occipitocervico-thoracic junction (Occiput-T3), these systems are indicated for: DDD (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies); spondylolisthesis; spinal stenosis; fracture/dislocation; atlantoaxial fracture with instability; occipitocervical dislocation; revision of previous cervical spine surgery; and tumors. The occipital bone screws are limited to occipital fixation only
FDA Recall Z-1691-2017 · Posted None
Advertisement
Recalling FirmZimmer Biomet Spine, Inc
Recall StatusTerminated
Product CodeMAX
Date Initiated2017-03-16
Date Terminated2018-07-30
Product Quantity141
Reason for Recall
Zimmer Biomet is recalling the Altius M-INI OCT Posterior Spinal Fixation System because the driver is unable to mate with the screw head tulip.
Firm Action
Zimmer Biomet sent an Urgent Medical Device Recall letter on March 20, 2017, to all affected consignees. The letter identified the product, the problem and the action to be taken by the consignee. Consignee's Responsibilities
1. Review this notification and ensure affected team members are aware of the contents.
2. Prior to sending Altius Kits to central processing, perform a functional check of the driver withall screws in the kit to ensure they mate. If the drivers do not mate, please return in
accordance with Step 3b.
3. Complete the Certification of Acknowledgement portion of Attachment 1 Inventory Return
Certification Form Phase I.
b. If the driver is unable to mate with the screws in the set, immediately sterilize and return the
affected nonconforming product from your distributorship and affected hospitals within your
territory along with a completed Attachment 1 Inventory Return Certification Form
Phase I and Attachment 2 Certificate of Sterilization to Zimmer Biomet.
i. For each return, send a copy of Attachment 1 to
[email protected]
ii. Include a hardcopy of Attachment 1 with your shipment for immediate
processing.
iii. Include a copy of Attachment 2 Certificate of Sterilization with returned
instruments.
iv. Mark the outside of the returns box(es) clearly with RECALL.
4. Distributors will notify their accounts to which the product was further distributed and to locate
and return products from these accounts. Using the Additional Accounts Form provided with
the email notice sent to your facility, return contact information for any additional hospitals that
have the affected product. If there are no additional users to notify, please confirm this by
annotating and returning the form provided indicating that no additional users have been
identified.
5. Retain a copy of your field action acknowledgement and product return forms for your records in
the event of a compliance audit of your facility.
6. If
Root Cause
Device Design
Distribution Pattern
Worldwide Distribution - US (nationwide) and Internationally to the Netherlands
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.