IMMULITE 2000/IMMULITE 2000 XPi BR-MA (CA15-3)

FDA Recall Z-1691-2018 · Posted 2018-05-04

Recall StatusTerminated
Product CodeMOI
Date Initiated2018-01-02
Date Terminated2023-02-10

Reason for Recall

Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.

Firm Action

Siemens Healthcare Diagnostics issued a recall in March 2017 for products with incorrect units for biotin in the Instructions for Use. After further investigation, recalling firm determined that more Siemens products also had missing or incorrect information in the IFUs for which two recalls were initiated, and Amended Corrections and Removals Reports were submitted to the FDA

Root Cause

Error in labeling

Distribution Pattern

Nationwide and Canada, Mexico, and Thailand

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.