Terminated

NeoBase Succinylacetone Assay Solution, Product Number: 3042-0020; Affected UDI Numbers: (01)06438147295708(17)180531(10)654184 (01)06438147295708(17)180531(10)653414 (01)06438147295708(17)180531(10)653657 (01)06438147295708(17)180630(10)654923 (01)06438147295708(17)180630(10)655680 (01)06438147295708(17)180630(10)655300 (01)06438147295708(17)180630(10)654924 (01)06438147295708(17)180831(10)656025 (01)06438147295708(17)180831(10)657055 (01)06438147295708(17)180831(10)656170 (01)06438147295708(17)180831(10)656320 (01)06438147295708(17)181031(10)657768 (01)06438147295708(17)181031(10)657513 (01)06438147295708(17)181031(10)658006

FDA Recall Z-1692-2018 · Posted None

Recall StatusTerminated
Product CodeNQL
Date Initiated2017-11-02
Date Terminated2019-09-17
Product Quantity5275

Reason for Recall

There is a potential for leaking vials for certain lot numbers where, if proper attire is not used as instructed, a transient minor allergic skin reaction may result when handling leaking succinylacetone reagent vials.

Firm Action

Actions to be taken: - Inspect your inventory of the NeoBase Succinylacetone Assay Solution vials for the affected lots. - If you notice leakage or deformed stopper seals, destroy the affected vials according to local instructions and regulations. - Complete the Response Form with the quantity and identity of affected Succinylacetone Assay Solution vials in your inventory. - Return the Response Form to PerkinElmer. - PerkinElmer will replace the quantity of leaking Succinylacetone Assay Solution vials found and destroyed from your inventory

Root Cause

Process control

Distribution Pattern

Distributed in 20 US states: AL, AR, AZ, CA, CO, FL, GA, IL, IN, KS, KY, LA, MI, MO, NV, OH, OR, SC, TN, TX.