Siemens Mammomat Inspiration - full-field digital mammography Product Usage: Mammography

FDA Recall Z-1694-2013 · Posted 2013-07-11

Recall StatusTerminated
Product CodeMUE
Date Initiated2013-02-28
Date Terminated2014-02-21
Product Quantity3

Reason for Recall

Firm became aware of an unintended behavior when using the Mammomat Inspiration. The possibility exists that a malfunction may occur during Stereo or Tomo acquisitions if the tomo clutch was incorrectly adjusted. Although the possibility is rare, the system could move while a patient's breast is compressed resulting in a possible bruise. Firm released Update Instructions XP031/12/S for this iss

Firm Action

Siemens issued an Update Instructions XP031/12/S by letter to the affected customers. For affected sites, a Siemens Service Engineer will inspect the unit and make any adjustments needed. Following the inspection and adjustments, a proper calibration will be completed. For questions call 610-448-3237 and 610-446-4634.

Root Cause

Device Design

Distribution Pattern

United States Nationwide Distribution in the states of ND, OH and SC.