SynchroMed II implantable drug infusion pump, Model 8637-40,

FDA Recall Z-1694-2017 · Posted None

Recall StatusTerminated
Product CodeLKK
Date Initiated2017-02-09
Date Terminated2018-09-04
Product Quantity1

Reason for Recall

Medtronic received a complaint that there was an error code displayed on the programmer when the physician attempted to interrogate an implanted SynchroMed II pump. The error code prevented the physician from updating the pump; however the pump was providing therapy.

Firm Action

A Medtronic representative visited the clinic on 2/9/2017 and corrected the SynchroMed II pump by resetting the pump memory using a SynchroMed Lab Programmer running on a Medtronic laptop with a validated protocol. Consignee was given a Medtronic "Urgent Medical Device Correction" letter dated February 8, 2017. The letter described the Issue and Actions to be taken by the consignee. For questions contact your Medtronic Representative Medtronic or Neuromodulation Technical Services at 1-800-707-0933.

Root Cause

Under Investigation by firm

Distribution Pattern

IL