SynchroMed II implantable drug infusion pump, Model 8637-40,
FDA Recall Z-1694-2017 · Posted None
Advertisement
Recalling FirmMedtronic Neuromodulation
Recall StatusTerminated
Product CodeLKK
Date Initiated2017-02-09
Date Terminated2018-09-04
Product Quantity1
Reason for Recall
Medtronic received a complaint that there was an error code
displayed on the programmer when the physician attempted
to interrogate an implanted SynchroMed II pump. The error
code prevented the physician from updating the pump;
however the pump was providing therapy.
Firm Action
A Medtronic representative visited the clinic on 2/9/2017 and corrected the SynchroMed II pump by resetting the pump memory using a SynchroMed Lab Programmer running on a Medtronic laptop with a validated protocol.
Consignee was given a Medtronic "Urgent Medical Device Correction" letter dated February 8, 2017. The letter described the Issue and Actions to be taken by the consignee. For questions contact your Medtronic Representative Medtronic or Neuromodulation Technical Services at 1-800-707-0933.
Root Cause
Under Investigation by firm
Distribution Pattern
IL
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.